Sch"after+Kirchhoff SK1024U3PD Bedienerhandbuch Seite 10

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 24
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 9
Page 10
3. Camera Control and Performing a Scan
Schäfter + Kirchhoff © 2014 • Line Scan Camera SK1024U3PD Manual (05.2014) • shared_CameraControl(2)_Adjustments_USB-mc.indd (05.2014)
3.2 Adjustments for Optimum Scan Results
Lens focussing
Sensor alignment
Gain/Offset
Shading correction
Integration time
Synchronization of the sensor exposure and
the object surface velocity, trigger mode
options
Prior to a scan, the following adjustments and parameter settings should be considered for optimum scan
signals:
Start with the signal window / oscilloscope display. Any changes in the optical system or camera parameters
are displayed in real-time when using an open dialog box.
Lens Focussing
To focus a line scan camera, the variations in edge steepness at dark/bright transitions and the modulation of the
line scan signal are desisive.
Adjust the focus using a fully opened aperture to restrict the depth of field and to amplify the effects of focus
adjustments.
A fully open aperture might cause a too high a signal amplitude, in which case the integration time should be
shortened, as described in Integration Time, p. 13.
Out of focus: edges are indistinct, signal
peaks blurred with low density modulation
Optimum focus: dark-bright transitions are
sharp edged, highly modulated signal
peaks with high frequency density
variations